Spectrum

ICP-OES Testing Services | Inductively Coupled Plasma Optical Emission Spectrometry for Elemental Analysis – Jituotech

Shanghai Jituotech Material Technology Co., Ltd. provides professional ICP-OES (Inductively Coupled Plasma Optical Emission Spectrometry) testing services using an Agilent 5110 ICP-OES instrument. Based on characteristic elemental emission spectra, ICP-OES enables qualitative and quantitative multi-element analysis of metals and selected non-metallic elements. Our services support elemental composition analysis, metal impurity testing, and material characterization for polymers, chemicals, ceramics, metals, composites, fine chemicals, and biomaterials, providing analytical data for research and development, formulation optimization, raw material evaluation, and quality control.

ICP测试-上海剂拓科技

ICP-OES Testing Services – Multi-Element Analysis, Metal Impurity Testing & Material Composition Characterization

Shanghai Jituotech Material Technology Co., Ltd. provides professional ICP-OES (Inductively Coupled Plasma Optical Emission Spectrometry) testing services for a wide range of material systems, including polymers, chemical materials, inorganic ceramics, metals, composites, fine chemicals, and biomaterials.

ICP-OES is an instrumental analytical technique used for qualitative and quantitative elemental analysis based on characteristic emission spectra. After appropriate sample preparation, the sample is introduced into a high-temperature inductively coupled plasma, where atoms and ions are excited and emit light at characteristic wavelengths. By measuring the wavelengths and intensities of these emission lines, the elements present in a sample can be identified and quantified using calibration standards. ICP-OES Testing Services Overview

ICP-OES is a widely used analytical technique for elemental composition analysis. Its main advantages include multi-element capability, relatively high analytical throughput, broad elemental coverage, and quantitative measurement capability.

Jituotech's ICP-OES testing services address the following analytical requirements:

Which elements are present in a material? Identify or screen selected metallic and non-metallic elements.

What is the concentration of a target element? Quantify elemental concentrations or mass fractions using calibrated standards.

Does the material contain metal impurities? Determine selected metallic impurities in raw materials, intermediates, or finished products.

Are elemental compositions consistent between batches? Compare element concentrations across different batches, suppliers, or formulations.

Does the material contain the expected inorganic components? Support the analysis of inorganic fillers, metallic components, catalyst residues, and functional elements.

Does the product meet specific elemental quality requirements? Perform quantitative analysis according to target elements, analytical limits, and applicable methods.

ICP-OES Service Information

Item

Service Information

Analytical technique

ICP-OES

Instrument model

Agilent 5110

Analytical principle

Characteristic elemental emission spectra

Analytical targets

Metals and selected non-metallic elements

Analytical capability

Qualitative and quantitative multi-element analysis

Sample forms

Solid and liquid

Sample quantity

100 mg–5 g as recommended on the original service page; actual requirements depend on sample type and preparation

Sample preparation

Dissolution, acid digestion, or other suitable pretreatment methods

Typical data output

Element identification, elemental concentrations, calibration curves, spectra, and analytical results

Typical applications

Raw material analysis, metal impurity testing, formulation research, and quality control What Is ICP-OES (Inductively Coupled Plasma Optical Emission Spectrometry)?

ICP-OES stands for Inductively Coupled Plasma Optical Emission Spectrometry. It is also commonly referred to as ICP-AES (Inductively Coupled Plasma Atomic Emission Spectrometry).

ICP-OES identifies and quantifies elements by measuring the characteristic light emitted by atoms or ions excited in an inductively coupled plasma.

Each element has characteristic emission lines. The instrument separates emitted light by wavelength and measures the intensity of selected spectral lines. Within an appropriate calibration range, emission intensity is related to the concentration of the target element.

How Does ICP-OES Work?

Sample Preparation: Solid or liquid samples are prepared and converted into a suitable solution for analysis.

Nebulization: The sample solution is converted into an aerosol through the sample introduction system.

Plasma Excitation: The aerosol enters a high-temperature argon plasma, where analytes are atomized, excited, and ionized.

Emission Generation: Excited atoms or ions emit light at characteristic wavelengths as they return to lower energy states.

Optical Detection: The optical system separates emission lines, and the detector records their intensities.

Quantitative Analysis: Calibration curves prepared from standard solutions are used to calculate elemental concentrations.

Result Calculation: Sample mass, digestion volume, dilution factors, and blank results are used to calculate elemental content in the original sample. Which Elements Can Be Analyzed by ICP-OES?

ICP-OES can be used to analyze a wide range of metallic elements and selected non-metallic elements.

Element Category

Representative Elements

Alkali metals

Li, Na, K

Alkaline earth metals

Mg, Ca, Sr, Ba

Transition metals

Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn

Precious metals and related elements

Ag, Au, Pt, Pd

Heavy metals and selected hazardous elements

Pb, Cd, Hg, As, Sb, Bi

Refractory metals

Mo, W, Ta, Nb, Zr, Hf

Rare earth elements

La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Yb, Lu, Y

Selected non-metallic elements

P, S, Si, B, Se, etc.

Note: These are representative examples of elements that may be analyzed by ICP-OES. Actual applicability depends on instrument configuration, spectral lines, sample matrix, concentration range, and method validation.

H2: What Can ICP-OES Measure? Typical Applications of ICP-OES Testing

ICP-OES is primarily used for qualitative and quantitative elemental analysis. It supports material composition characterization, metal impurity testing, inorganic component analysis, and selected elemental quality-control applications.

  1. Metal Element Content Analysis

Determine the concentrations of one or multiple metallic elements to support material composition analysis and quality control.

Typical applications:

Elemental composition analysis of metals and alloys;

Metal content determination in inorganic powders;

Elemental analysis of chemical raw materials;

Residual metal analysis in polymer materials;

Metallic component analysis in composite materials.

  1. Trace and Low-Level Metal Impurity Analysis Determine selected metal impurities in materials to support raw material purity evaluation and product quality control.

Typical applications:

Metal impurity testing in high-purity chemicals;

Catalyst residue analysis in fine chemicals;

Residual metal catalyst analysis in polymers;

Impurity element analysis in ceramic powders and functional materials;

Selected metal element analysis in biomaterials.

Note: Trace analysis capability depends on the target element, sample matrix, and sample preparation. For extremely low concentrations or ultra-trace analysis, ICP-MS may be more appropriate.

  1. Inorganic Filler and Functional Element Analysis

Determine selected elemental concentrations to support the characterization of inorganic fillers, functional components, and material formulations.

Typical applications:

Elemental analysis of inorganic fillers in plastics and rubber;

Metallic component analysis in polymer composites;

Elemental composition studies of ceramic materials;

Inorganic element analysis of coatings;

Functional powder and additive characterization. 4. Heavy Metal and Selected Hazardous Element Testing

Determine selected elements such as lead, cadmium, arsenic, chromium, and nickel to support material quality control and relevant compliance assessments.

Typical applications:

Selected heavy metal analysis in chemical raw materials;

Hazardous element analysis in inorganic powders;

Metal element testing in plastics and rubber;

Target element analysis in ceramics and composites;

Raw material supplier quality evaluation. 5. Raw Material and Batch Consistency Analysis

Compare elemental concentrations across multiple batches using consistent sample preparation and analytical conditions.

Typical applications:

Raw material supplier screening;

Elemental composition comparison between product batches;

Manufacturing process consistency evaluation;

Formulation comparison before and after adjustment;

QA/QC quality control.

  1. Catalyst and Functional Material Analysis

Determine active metals, promoters, and other target elements in catalysts and functional materials.

Typical applications:

Metal loading analysis in catalysts;

Precious metal and transition metal determination;

Dopant analysis in functional ceramics;

Metal element analysis in battery-related materials;

Elemental composition evaluation of advanced functional materials.

  1. Elemental Mass Balance and Process Research

Compare elemental concentrations in raw materials, intermediates, and finished products to investigate elemental introduction, loss, or residue during processing.

Typical applications:

Material preparation process optimization;

Metal recovery studies;

Catalyst elemental analysis before and after use;

Elemental mass balance between raw materials and final products;

Material purification and impurity removal evaluation. H2: What Materials Can Be Tested by ICP-OES?

ICP-OES is suitable for a wide range of materials that can be converted into stable solutions through appropriate sample preparation.

Jituotech's ICP-OES testing services cover the following material categories:

Material Category

Typical Analytical Applications

Polymers

Residual metal catalysts, inorganic elements, additive-related elements

Plastics and rubber

Metal elements, inorganic filler-related elements, heavy metals

Chemical materials

Raw material composition, metal impurities, catalyst residues

Inorganic ceramics

Major metal elements, dopants, impurity elements

Metals and alloys

Multi-element composition, alloying elements, impurities

Composite materials

Matrix- and filler-related elements, metallic components

Fine chemicals

Metal impurities, catalyst residues, target element concentrations

Biomaterials

Selected metals and inorganic elements

Functional powders

Elemental composition, dopant elements, impurities

Catalysts and additives

Active metals, promoters, and other target elements

How Does Sample Preparation Differ Between Materials?

ICP-OES generally requires samples to be introduced as suitable solutions. Different material systems require different sample preparation methods.

Sample Type

Typical Sample Preparation Approach

Water-soluble chemicals

Appropriate dissolution, dilution, and filtration where necessary

Metals and alloys

Acid digestion using an acid system suitable for the alloy composition

Ceramics and inorganic powders

Acid digestion or another suitable pretreatment depending on acid resistance

Plastics and rubber

Usually require digestion or ashing before solution analysis

Polymers and composites

Digestion procedures developed according to polymer matrices and fillers

Fine chemicals

Dissolution or digestion based on solubility and matrix interference

Biomaterials

Appropriate digestion based on organic matrix composition and target elements

Which Samples May Require Special Assessment Before ICP-OES Testing?

The following samples require a feasibility and safety assessment:

Samples that cannot be effectively dissolved or digested using available preparation methods;

Samples containing large quantities of insoluble particles that prevent representative solution preparation;

Samples that may cause severe matrix interference or instrument contamination;

Samples containing high concentrations of acids, salts, or organic solvents beyond the compatible range of the sample introduction system;

Samples that may release toxic, flammable, or corrosive gases during digestion;

Samples requiring chemical speciation rather than total elemental analysis.

For refractory ceramics, silicates, precious metal alloys, and specialized composites, please provide sample composition information before submission so that an appropriate digestion strategy can be evaluated.

H2: ICP-OES Instrument Model

Jituotech's ICP-OES testing service uses an Agilent 5110 Inductively Coupled Plasma Optical Emission Spectrometer.

Instrument Brand: Agilent

Instrument Model: Agilent 5110 ICP-OES

Instrument Type: Inductively Coupled Plasma Optical Emission Spectrometer

The Agilent 5110 ICP-OES excites elements in a high-temperature argon plasma and detects their characteristic emission lines. It is used for qualitative and quantitative multi-element analysis.

Depending on the instrument configuration, the Agilent 5110 series supports different plasma viewing configurations, including radial and vertical dual-view configurations. The actual viewing mode should be confirmed according to the laboratory's installed equipment.

Image suggestion:

H2: ICP-OES Instrument Technical Specifications

The following information is compiled from Jituotech's existing ICP-OES service page and Agilent 5110 technical documentation.

Technical Parameter

Published Information or Reference

Instrument brand

Agilent

Instrument model

5110 ICP-OES

Analytical technique

Inductively coupled plasma optical emission spectrometry

Detector

VistaChip II CCD (model technical documentation)

Wavelength range

167–785 nm (model technical documentation)

Signal stability

Typically less than 1% RSD over 8 hours without internal standardization or drift correction

Sample type

Liquid samples and solid samples after appropriate preparation

Analytical capability

Qualitative and quantitative multi-element analysis

Typical Analytical Wavelengths and Spectral Resolution

Representative wavelengths and spectral resolution values for the Agilent 5110 are listed below.

Element

Symbol

Wavelength (nm)

Typical Resolution FWHM (pm)

Arsenic

As

188.980

<7

Molybdenum

Mo

202.032

<7

Zinc

Zn

213.857

<7.5

Lead

Pb

220.353

<8

Chromium

Cr

267.716

<9.5

Copper

Cu

327.395

<13

Barium

Ba

614.171

<34

Notes:

These wavelengths and resolution values are representative technical specifications for the Agilent 5110.

Actual analytical wavelengths should be selected according to emission-line characteristics, sample matrix, spectral interference, and concentration range.

Spectral resolution is not equivalent to an elemental detection limit or quantification limit.

The "1 ppm" parameter mentioned on the original service page should be technically verified before publication. It should not be presented as a universal detection limit for all elements.

Actual analytical performance depends on instrument configuration, sample preparation, method parameters, and quality-control results.

H2: ICP-OES Sample Requirements

Appropriate sample preparation helps ensure representative results, analytical accuracy, and testing safety.

Item

Sample Requirements

Sample form

Solid or liquid

Sample quantity

100 mg–5 g as recommended on the original service page; confirm according to material type and preparation method

Sample homogeneity

Samples should be representative and well mixed; powders and particles should be homogenized where appropriate

Packaging

Use clean, sealed containers compatible with the sample

Solubility

The original service page recommends complete dissolution in suitable inorganic solvents or acid systems

Sample information

Provide sample name, major components, target elements, and expected concentration range

Safety information

Disclose corrosive, toxic, flammable, or other hazardous properties

What Information Should Be Provided Before ICP-OES Testing?

To establish an appropriate testing program, customers are encouraged to provide:

Sample name and material category;

Main chemical components or known formulation;

Target elements to be analyzed;

Expected concentration range of each target element;

Sample solubility and known acid resistance;

Required testing standards, if applicable;

Whether a formal report, raw data, or spectral plots are required;

Whether the results are intended for R&D comparison, quality control, or compliance assessment.

Can Solid Samples Be Tested Directly by ICP-OES?

Ordinary solid samples are generally not introduced directly into conventional solution-based ICP-OES systems without appropriate pretreatment.

Solid samples usually require dissolution, acid digestion, microwave digestion, fusion, or another suitable preparation method to transfer target elements into a compatible solution.

Refractory ceramics, silicates, and specialized metal materials may require specific digestion procedures. Incomplete digestion can result in underestimated elemental concentrations.

H2: ICP-OES Testing Process

Jituotech's ICP-OES testing process includes sample information review, preparation method development, instrumental analysis, data calculation, and report review.

  1. Testing Requirement Consultation

Confirm sample category, target elements, expected concentration range, testing objectives, and reporting requirements.

  1. Sample Assessment

Evaluate sample homogeneity, solubility, matrix composition, and safety to determine whether ICP-OES analysis is suitable.

  1. Sample Preparation

Select an appropriate dissolution, acid digestion, microwave digestion, or other preparation method to transfer target elements into a suitable solution.

  1. Standard Solution and Calibration Curve Preparation

Prepare elemental standards within the appropriate concentration range and establish calibration curves. Blanks and quality-control samples are included where required.

  1. Instrumental Analysis

Introduce the prepared sample solution into the ICP-OES instrument and measure the characteristic emission lines of the target elements.

  1. Data Analysis and Result Calculation

Calculate elemental concentrations using calibration data, sample mass, digestion volume, dilution factors, and blank results.

  1. Data Review and Report Delivery

Review analytical results and quality-control data, then deliver elemental concentrations, relevant spectra, and testing reports according to the agreed project requirements.

ICP-OES Testing Workflow Diagram

flowchart TD A[Testing Requirement Consultation] --> B[Sample and Target Element Confirmation] B --> C[Sample Composition and Safety Assessment] C --> D{Suitable for ICP-OES?} D -->|No| E[Evaluate Alternative Preparation or Analytical Method] E --> B D -->|Yes| F[Develop Sample Preparation Method] F --> G[Sample Dissolution or Acid Digestion] G --> H[Standard Solution and Calibration Curve Preparation] H --> I[ICP-OES Instrumental Analysis] I --> J[Emission Spectrum Acquisition] J --> K[Elemental Concentration Calculation] K --> L[Quality Control and Data Review] L --> M[Testing Report Delivery]

H2: ICP-OES Testing Standards

ICP-OES testing standards should be selected according to sample type, target elements, and analytical objectives. Sample digestion, instrument conditions, and calculation procedures may differ between material systems.

The following are relevant reference standards and methods:

Standard

Standard Title

Typical Application

ISO 19050:2021

Rubber, raw, vulcanized — Determination of metal content by ICP-OES

Determination of major and trace metal contents in raw and vulcanized rubber

ISO 11885:2007

Water quality — Determination of selected elements by ICP-OES

Determination of selected elements in water

EPA SW-846 Method 6010D

Inductively Coupled Plasma-Optical Emission Spectrometry

Multi-element analysis of appropriately prepared samples

ASTM D5185

Standard Test Method for Multielement Determination of Used and Unused Lubricating Oils and Base Oils by Inductively Coupled Plasma Atomic Emission Spectrometry

Multi-element analysis of lubricating oils and base oils; confirm the current edition and scope before use

Standard Applicability Notes

ISO 19050:2021: Covers determination of major and trace metal contents in raw and vulcanized rubber by ICP-OES.

ISO 11885:2007: Primarily addresses selected elements in different types of water and also covers digested sludge and sediment samples.

EPA Method 6010D: Describes multi-element analysis by ICP-OES. Sample preparation should be appropriate for the target elements and sample matrix.

ASTM D5185: Addresses multi-element analysis of lubricating oils and base oils. Specific sample scope and analytes should be confirmed against the standard text.

Note: These standards are references and should not be applied indiscriminately to all materials. Before testing, confirm sample type, standard scope, preparation conditions, target analytes, and laboratory capability.

Important: ISO 22125-2 is related to ICP-MS and should not be listed as an ICP-OES testing standard.

H2: ICP-OES vs. Other Elemental Analysis Techniques

ICP-OES, ICP-MS, AAS, and XRF are all used for elemental analysis, but they differ in analytical principles, sensitivity, sample preparation requirements, and application scope.

Comparison of ICP-OES, ICP-MS, AAS, and XRF

Technique

Full Name

Analytical Principle

Main Advantages

Typical Applications

ICP-OES

Inductively Coupled Plasma Optical Emission Spectrometry

Measures characteristic light emitted by excited elements in plasma

Multi-element analysis, relatively high throughput, broad elemental coverage

Metal content, material composition, impurity analysis

ICP-MS

Inductively Coupled Plasma Mass Spectrometry

Measures ions according to their mass-to-charge ratio

Generally lower detection limits, suitable for ultra-trace and isotope analysis

Ultra-trace metals and isotope analysis

AAS

Atomic Absorption Spectrometry

Measures absorption of characteristic light by ground-state atoms

Established analytical technique for selected elements

Metal and selected heavy metal analysis

XRF

X-ray Fluorescence Spectrometry

Measures characteristic fluorescent X-rays emitted after X-ray excitation

Rapid solid-sample screening; non-destructive testing in suitable applications

Metals, minerals, ceramics, and solid material screening

What Is the Difference Between ICP-OES and ICP-MS?

ICP-OES measures characteristic light emitted by excited elements, while ICP-MS detects ions according to their mass-to-charge ratios.

In general:

ICP-OES is suitable for multi-element analysis of major, minor, and selected trace elements.

ICP-MS is generally more suitable for ultra-trace elemental analysis.

ICP-OES can be a practical choice when target element concentrations are relatively high.

ICP-MS or another appropriate technique should be considered for isotope ratio and isotope composition studies.

The final selection should consider target elements, expected concentrations, sample matrix, detection limits, and project requirements.

What Is the Difference Between ICP-OES and XRF?

ICP-OES generally requires samples to be dissolved or digested before solution analysis. XRF can analyze suitable solid samples directly.

ICP-OES: Suitable for quantitative multi-element analysis of prepared solutions. Results depend on digestion efficiency and solution matrix effects.

XRF: Suitable for rapid elemental screening and selected quantitative analysis of solid materials, often without complete sample digestion.

ICP-OES may be preferred when solution-based quantitative results are required, while XRF can be considered for rapid elemental screening of solid materials.

H2: Three Typical ICP-OES Testing Case Studies

The following are representative application examples developed to illustrate ICP-OES testing strategies, sample preparation, and data interpretation. They are not disclosed Jituotech client projects or actual customer test reports. For final website publication, they should be replaced with anonymized real projects and verified analytical data.

Case Study 1: Elemental Composition and Impurity Analysis of Inorganic Ceramic Powders

Test Sample:

Oxide ceramic powders or functional inorganic powders.

Testing Objectives:

Determine major metallic element concentrations;

Analyze trace metal impurities in raw materials;

Compare elemental composition between different powder batches;

Support ceramic formulation development and raw material quality control.

Testing Method:

Select an appropriate pretreatment method according to the sample composition and acid solubility.

Weigh and digest the sample to transfer target elements into solution.

Prepare elemental standard solutions and establish calibration curves.

Measure characteristic emission lines using ICP-OES.

Calculate elemental concentrations based on sample mass, digestion volume, and dilution factors.

Data and Spectra Obtained:

Characteristic emission spectra of target elements;

Calibration curves and related calibration data;

Concentrations of target elements;

Calculated elemental mass fractions in the original sample;

Elemental composition comparison between batches.